TURBIDITY METER, TURBIDIMETER, PORTABLE TURBIDITY METER أجهزة قياس عكارة المياة
Portable Turbidity Meter AMT21          
Features:
1. Fast and accurate measurements.
2. Measurement method meet ISO 7027
3. Large 45x25mm LCD display
4. Unit NTU, FTU, EBC changeable
5. Memory function stores and recalls up to 150 points
6. Low battery alarm
7. Auto off after 10 minutes
Specifications:
1. Range: 0~19.99NTU, 20~199.9NTU, 200~1000NTU
2. Resolution: 0.01, 0.1, 1NTU
3. Accuracy: ±3% <500NTU; ±5% >500NTU
4. Response time: <10s
5. Operating: 0~50°C, <85% RH
6. Sample: 10ml
7. Data memory: 150 points
8. Power: 4 x AAA battery
9. Size: 70 x 135 x 65mm
10. Weight: 168g
 
Portable Turbidity Meter AMT27          
Features:
1. Easy-to-use portable turbidity meter is equipped with a standard carrying case.
2. 2 to 5 points calibration using the Formazin Standards.
3. Selectable four turbidity units, including the NTU (Nephelometric Turbidity Unit), FNU (Formazine Turbidity Unit), EBC (European Brewery Convention) and ASBC (American
Society of Brewing Chemists).
4. Single measurement mode automatically recognizes the stable reading.
5. Continuous measurement mode that can used for indexing and matching the sample vials.
6. Auto-Power Off helps user effectively conserve battery life.
7. System menu allows setting the 7 parameters, including the date and time, resolution, etc.
8. Reset feature automatically resumes all settings back to factory default settings.
9. Expanded memory stores and recalls up to 100 readings.
10. Stored data can be transferred into computer by USB communication interface.
11. Multi-mode power scheme (battery, power adapter, USB port) ensures that use the meter smoothly.
Specifications:
1. Measurement Method: ISO 7027 nephelometric method (90°)
2. Measurement Range: 0~1100 NTU/FNU, 0~275 EBC, 0~9999 ASBC
3. Resolution: 0.01 (0~100 NTU), 0.1 (100~999 NTU), 1 (999~1100 NTU)
4. Accuracy: ±2% (0~500 NTU), ±3% (501~1100 NTU)
5. Calibration Standards: 0.02, 10, 200, 500, 1000 NTU or User-Defined
6. Light Source: Infrared-emitting diode (850 nm wavelength)
7. Detector: Silicon Photodiode
8. Stray Light: <0.02 NTU
10. Sample Vials: 60(H) x 25(Dia.)mm
11. Power Off: Manual or Automatic (15 minutes)
12. Memory: Stores up to 100 data sets
13. Output: USB Communication Interface
14. Operating Temperature: 0~60°C
15. Power Requirements: 1 x 9V Battery
16. Dimensions: 180(L) x 85(W) x 70(H)mm
17. Weight: 300g

 

 
Portable Turbidity Meter TU001, TU002, TU003, TU004           

Summary of functions

Our scattered light turbidity meter is used for measuring the scattering degree of the lights generated by the insoluble grain matters suspended in water or transparent liquids, with capability to characterize the content of these suspended grain matters. It can be broadly applied in the measurement of turbidity in power plants, waterworks, treatment stations of domestic sewage, beverage factories, environmental protection departments, industrial water, wine and pharmacy industries, epidemic prevention divisions, hospitals, and etc.

Product Characteristics
1. AC & DC power supply, low power consumption, cabinet and handsome contour design, facilitating field and lab use
2. Configured with auto power-off function and low power dissipation circuit, able to effectively prolong service life of battery
3. Microcomputer, touch keyboard, backlight LCD, standard parallel RS232 data communication interface
4. Custom-made high strength long service life lamp-house, provided with data storage and inquiry function, can meet GLP requirements
5. Data nonlinear processing and data smoothing function, quick and automatic multi-points calibration, self-diagnostic information prompt, selectable span automatic or manual switch
6. Adopting big-screen highly transparent LCD display, with comfortable reading, and free of ill influence from sunlight
7. Adopting low-drift and high-accuracy integrated circuit, efficient and time proof light source, effectively ensuring long reliable performance
8. Unique locative structure and high-accuracy light path system, effectively ensuring exactness and repeatability
9. Low voltage indication, cuing users to change battery, to ensure exact and effective measurement
10. Configured with output port of measured values, possible to connect with recording meter

Specifications

ModelTU001TU002TU003TU004
Minimum Principle

90°scattered light

Minimum readout (NTU)0.010.0010.010.001
Measuring range (NTU)

0~50

0~10

0~100~10

0~500

0~1000~1000~100

 

0~10000~10000~500

 

  0~1000
   0~4000
Basic error (F.S.)±2%
Repeatability≤1%
Zero draft (F.S.)±0.5% (0.1 NTU)

±0.5% (0.02 NTU)

 
Portable Turbidity Meter TU005, TU006, TU007           
Specifications

Model

TU005

TU006TU007
Minimum Principle90° scattered light
Minimum readout (NTU)0.10.010.01
Measuring range (NTU)0~200 NTU0~20 NTU0~20, 0~200 NTU
Basic error F.S±8% (±2.5%F.S)
Repeatability≤1%
Zero draft (NTU)

±1%F.S

Power supply

DC1.5V x 5 AA alkaline dry cells; AC 220V/50Hz DC7.5V 0.2A adapter

External dimension235×75×65 mm
OthersLCD display, LED Light source, auto power off.
 
Benchtop Turbidity Meter TB200           
Features:
1. Standard benchtop turbidity meter is equipped with a 4.5" LCD display.
2. 2 to 7 points calibration using the Formazin Standards.
3. Selectable four turbidity units, including the NTU, FNU, EBC and ASBC.
4. TSS conversion factor ensures accurate measurement for Total Suspended Solids.
5. Continuous measurement mode that can used for indexing and matching the sample vials.
6. Hold function freezes current reading for easy viewing and recording.
7. Password protection prevents the unauthorized calibration and settings.
8. Help message as a operational guide that helps user quickly begin using the meter.
9. System menu allows setting the 9 parameters, including the date and time, resolution, etc.
10. Reset feature automatically resumes all settings back to factory default settings.
11. Expanded memory stores and recalls up to 200 readings.
12. Stored data can be transferred into computer by USB communication interface.
Specifications:
1. Measurement Method: ISO 7027 nephelometric method (90°)
2. Turbidity Range: 0~2000 NTU/FNU, 0~500 EBC, 0~9999 ASBC
3. TSS Range: Depends on conversion factor
4. Resolution: 0.01 (0~100 NTU), 0.1 (100~999 NTU), 1 (999~2000 NTU)
5. Accuracy: ±2% (0~500 NTU), ±3% (501~2000 NTU)
6. Calibration Standards: 0.02, 10, 200, 500, 1000, 1500, 2000 NTU or User-Defined
7. Light Source: Infrared-emitting diode (850 nm wavelength)
8. Detector: Silicon Photodiode
9. Stray Light: <0.02 NTU
10. Sample Vials: 60(H) x 25(Dia)mm
11. Power Off: Manual or Automatic (2 hours after last key pressed)
12. Reset Function: Yes
13. Memory: Stores up to 200 data sets
14. Output: USB Communication Interface
15. Operating Temperature: 0~60°C
16. Power Requirements: DC12V Power Adapter
17. Dimensions: 250(L) x 177(W) x 96(H)mm
18. Weight: 1.2kg
 
Bench Turbidity Meter TU008, TU009, TU010           

Summary of functions

TU serials scattered light turbidity meter is used for measuring the scattering degree of the lights generated by the insoluble grain matters suspended in water or transparent liquids, with capability to characterize the content of these suspended grain matters.

It can be broadly applied in the measurement of turbidity in power plants, waterworks, treatment stations of domestic sewage, beverage factories, environmental protection departments, industrial water, wine and pharmacy industries, epidemic prevention divisions, hospitals, and etc.

Product Characteristics

a. The streamlined shape and design, Easy operation with economy price.

b. Large LCD screen with clearly display. Stability electrical system with the low drift, high precision

c. Reliable location structure and high precision light system to make the instrument can work stability.

d. Color compensation ,could avoid the disturbing from the specimen, that can reflect the concept of turbidity correctly.

e. The high intensity of light, without change frequency, warm -up time 30 seconds to work fast.

Specifications

Model

TU008

TU009

TU010

Minimum Principle

90° scattered light

Minimum readout (NTU)0.010.10.01
Measuring range (NTU)

0~20 NTU

0~200NTU

0~20, 0~200 NTU

Basic error F.S

±8% (±2.5%F.S)

Repeatability

≤0.8%

Zero draft

±0.8%F.S

Power supply

AC 220V/50Hz, 3W

Others

Economic Type, Easy Operation, Good Stability.

 
Bench Turbidity Meter TU011, TU012, TU013           
Product Characteristics
a. The streamlined shape and design, Easy operation with economy price.
b. Large LCD screen with clearly display. Stability electrical system with the low drift, high precision
c. Reliable location structure and high precision light system to make the instrument can work stability.
d. Color compensation ,could avoid the disturbing from the specimen, that can reflect the concept of turbidity correctly.
e. The high intensity of light, without change frequency, warm -up time 30 seconds to work fast.
Specifications:
ModelTU011TU012TU013

Minimum Principle

90°Scattering Light
Minimum readout (NTU)0.1

0.01

0.1

Measuring range (NTU)

0~2000~20, 0~200

0~200, 0~800

Basic error (F.S)

±6% (±2.5%F.S)

Repeatability≤0.5%
Zero draft (F.S)±0.5% F.S.
CharacteristicsGeneral-purposeFor low measuring rangeFor high measuring range
 
Bench Turbidity Meter                                                     

Summary of Functions

TU Series scattered light turbidity meter is used for measuring the scattering degree of the lights generated by the insoluble grain matters suspended in water or transparent liquids, with capability to characterize the content of these suspended grain matters.

It can be broadly applied in the measurement of turbidity in power plants, waterworks, treatment stations of domestic sewage, beverage factories, environmental protection departments, industrial water, wine and pharmacy industries, epidemic prevention divisions, hospitals, and etc.

Product Characteristics

a. Adopting low-drift and high-accuracy integrated circuit, efficient and time proof light source, effectively

ensuring long reliable performance

b. Unique locative structure and high-accuracy light path system, effectively ensuring exactness and repeatability;

c. The configuration with the high-automatic & multi-function microcomputer, equipped with the RS232 serial communication interface to connect the printer.

d. The imbedded clock memory & storage system can store the measurement & correction data in real time, and can store and transfer 20 groups of the measurement data for long time, and can be used for the historic reference.

e. The quick automatic correction, with the prompt of the self-diagnosis information, rapid and stable response, reliable maintenance-free operation

Model (P with printer)

TU015

TU016

TU017

TU018

TU019

TU020

TU015PTU016PTU017PTU018PTU019PTU020P

Minimum Principle

90° scattered light

Minimum readout (NTU)

0.01

0.001

0.001

0.01

0.01

0.001

Measuring range (NTU)

0~20

0~10

0~10

0~50

0~10

0~10

0~200

0~100

0~100

0~500

0~100

0~100

----

----

0~500

----

0~500

0~500

----

----

----

----

0~1000

0~1000

Basic error (F.S)

±0.6% (±2% F.S)

Repeatability

≤0.5% F.S

Zero draft (F.S)

±0.5% F.S

NoteAverage measuring method, Year/Month/Day/Time display, Data record and auto range change, auto zero, 1~5 point calibration, RS232 port
 
Lab Grade Bench Turbidity Meter                                                     

Summary of functions

TU Series scattered light turbidity meter is used for measuring the scattering degree of the lights generated by the insoluble grain matters suspended in water or transparent liquids, with capability to characterize the content of these suspended grain matters.

It can be broadly applied in the measurement of turbidity in power plants, waterworks, treatment stations of domestic sewage, beverage factories, environmental protection departments, industrial water, wine and pharmacy industries, epidemic prevention divisions, hospitals, and etc.

Product Characteristics

1. Microcomputer, touch keyboard, backlight LCD, standard parallel RS232 data communication interface

2. Custom-made high strength long service life lamp-house, provided with data storage and inquiry function, can meet GLP requirements

3. Data nonlinear processing and data smoothing function, quick and automatic multi-points calibration, self-diagm,nostic information prompt, selectable span ¨C automatic or manual switch

4. Set average measuring mode with shortcut to obtain correct data within quickest time, applicable particularly to measure extreme-low turbidity and can be used to measure unstable water sample

5. Precise light route system, reliable positioning structure, effective tone compensation, direct turbidity reading , built-in or external printer is optional

Model (P with printer)

TU021

TU022

TU023

TU024

TU021PTU022PTU023PTU024P

Minimum Principle

90° scattered light

Minimum readout (NTU)

0.001

0.01

0.001

0.01

Measuring range (NTU)

0~10

0~10

0~10

0~50

0~100

0~100

0~100

0~500

0~1000

0~1000

0~1000

0~1000

0~2000

0~2000

0~4000

0~4000

Basic error (F.S)

±6% (±2% F.S, ±3% F.S, ±5% F.S)

Repeatability

≤0.5% F.S

Zero draft (F.S)

±0.5% F.S

NoteAverage measuring method, Year/Month/Day/Time display, Data record and auto range change, auto zero, 1~5 point calibration, RS232 port
 
Bench Turbidity Meter WGZ-200           
Features
1. It offers measurement of light scatter and decay of the insoluble particles suspended in water or transparent liquids.
2. This turbidity meter can be used for quantitative analysis of particle contents.
Technical Specifications
Light source: Tungsten halogen lamp 6V, 10W
Receiving element: Silicon photocell
Measurement range NTU: 0.00-50.0; 50.1-200 (automatic range switchover)
Readings display: 4 digits LED display
Allowable error of indication: ≤±8%
Zero Drift: ≤±3%FS
Stability of indication: ≤±1%FS
Repeatability: ≤2%
Sample bottle: Φ 25 mm×95 mm
Sample volume: 20 ~ 30 ml
Weight (Gross): 9kg
Overall dimensions: 460 x 430 x 250mm
 
Bench Turbidity Meter WGZ-2000          
Features
1. The WGZ-2000 Ratio turbidity meter is used for measurement of light scatter and decay of the insoluble particles suspended in water or transparent liquids.
2. If offers quantitative analysis of particle contents.
Technical Specifications
Light source: Tungsten halogen lamp 6V, 10W
Receiving element: Silicon photocell
Measurement range NTU: 0.00-50.0; 50.1-200; 201-2000 (automatic range switchover)
Readings display: 4 digits LED display
Allowable error of indication:
a. 0-200NTU: ≤±8%
b. 201-2000NTU: ≤±6%
Zero Drift: ≤±3%FS
Stability of indication: ≤±1%FS
Repeatability: ≤2%
Sample bottle: Φ 25 mm×95 mm
Sample volume: 20 ~ 30ml
Weight (Gross): 9kg
Overall dimensions: 460 x 430 x 250mm
 
Turbidity Meter A005-6          
Specifications 1:

Size: Diameter 60mm* Length 256mm
Weight:1.65 KG
Main Materials:Main:Body:SUS316L (Ordinary Version),Titanium Alloy (Seawater Version) Upper and Lower Cover: PVC; Cable: PVC
Waterproof Rate:IP68/NEMA6P
Measurement Range:0.01-100 NTU 、0.01-4000 NTU
Indication Resolution: Less than ± 2% of the measured value, or ± 0.1 NTU Maximax criterion
Pressure Range:≤0.4Mpa
Flow velocity:≤2.5m/s、8.2ft/s
Storage Temperature:-15~65℃
Environment Temperature:0~45℃
Calibration:Sample Calibration, Slope Calibration
Cable Length:Standard 10-Meter Cable, Max Length: 100 Meters
Warranty Period:1 Year
High Voltage Baffle:Aviation Connector, Cable Connector


Specifications 2:

Size:145*125*162mm L*W*H
Weight:1.3KG
Material of outside shell: lower casing: Aluminum with powder coveringcover:PA66+GF25+FR
Waterproof Rate:IP65/NEMA4X
Storage Temperature:-20 to 70℃
Operation Temperature:-15 to 60℃
Power Supply:AC:AC220V, 50HZ, 5W DC:DC24V
Output: three-way analog output 4-20mA, response parameters and corresponding scope can be programmed
Note: the maximum load is 500 ohms
Relay: three-way relay can be set up, and response parameters and response values can be programmed
Display Output:128 * 64 dot matrix LCD with LED backlight, which can be operated under the direct sunlight
Digital Communication: MODBUS RS485 communication function, which can transmit real-time measurements
Warranty Period:1 year

 